back to HUNG CHANG

 

Protek-Z9216

 

Digital LCR Meter

Features

  • 0.05% Basic Accuracy
  • 5 digit display
  • Test frequencies from 100Hz to 100KHz
  • 2V DC internal bias and external bias to 40V DC
  • RS-232C and GPIB(option) Interface

Specification

  • Display
    - Mode : Auto, R+Q, L+Q, C+D, C+R
    - Addition Circuit : Series, Parallel
    - Display Variables : Measurement Value, Diversion, % Diversion,
    Bin Number Diversion and % Diversion are calculated values.
    - Average Measurement Range :
    Average values after 2 ~ 10 times measurement
    R+Q : R, 0.0001 Ohm~ 2000M Ohm Q, 0.00001 ~ 50
    L+Q : L, 0.00001uH ~ 99999H Q, 0.00001 ~ 50
    C+D : C, 0.0001pF ~ 99999uF D, 0.000001 ~ 10
    C+R : C, 0.001pF ~ 99999uF R, 0.00001 ~ 99999K Ohm

  • Measurement Condition
    Frequency : 100Hz, 120Hz, 1KHz, 10KHz, 100KHz Accuracy : ±100ppm
    Voltage : 0.1V, 0.25V, 1Vrms
    Speed : Slow, Medium, Fast
    Frequency    |    Slow    |     Medium  |     Fast   
     100KHz           2.8/sec        14/sec         28/sec
     10KHz             2.8/sec        14/sec         27/sec
     1KHz               2.7/sec        13/sec         24/sec
     120Hz             0.7/sec        2.8/sec          7/sec
     100Hz             0.6/sec        2.4/sec          6/sec

    Range Selection : Automatic and Manual
    Circuit : Series and Parallel Circuit
    Triggering : Continuous, Manual or RS232C, GPIB Handler External Trigger via Interface
    Bias Voltage : Internal 2.0 VDC ±2% External 0 ~ +40VDC(fused @0.25A)
    Connection : 4 Circuits BNC Connection
    Protection Circuit : Built-in Protection Circuit when charged condenser measurement (1 time)
    Null Calibration :Open and Short Correction and Compensation

  • Operation
    0~50°C, below RH 80% 30 min. preheating 23°C±5°C

  • Operation Voltage
    100V, 120V, 220V, 240V ±10% 50/60Hz

  • Dimension & Weight
    363(W) x 109(H) x 386(D)mm, 5.8kg

  • Option
    GPIB Handler Interface : IEEE-488 Interface and Handler Interface
    Kelvin Clips : When connection with different Fixture
    MD Tweezers : Surface Mount Fixture Connection
    Radial Fixture : Radial or Axial Type Fixture Measurement (Direct
    Connection to LCR Meter)

  • Basic Accessories
    - Power Cord 1EA
    - Fuse 2EA
    - Manual 1EA
  • Radial Fixture
    Radial or Axial Type Fixture Measurment (Direct Connection to LCR Meter)

 


  Back to HUNG CHANG Products or
  Back to top of the page

For further information, please
Contact Us

Request a Quote

 


GIGA Electronics Ltd. - Innovative Electronic Solutions

At GIGA Electronics Ltd. our main aim is to offer to our customers the best and most effective solution to meet their
test & measurement needs through exceptional quality and reliability of our products and services.
We take a personal approach to each customer according to his individual needs.