back to HUNG CHANG

 

Protek-608

 

Digital Multimeter

Features

  • 4 4/5 Digit, Dual Display
  • 50,000 Count True RMS
  • True RMS for AC or AC+DC
  • Accuracy : 0.05%(DC Volt)
  • Auto Calibration
  • PC Interface with RS232C
  • LCD Backlight
  • New style of bar graph display
    based on binary bit expression method
  • ±Peak detection mode for high speed signal capture
  • Easy determination on with GO/NG mode
  • Low battery indication

Specification

  • DC mV & V

Function

Range

Accuracy

Input Impedance

Protection

DC mV

500.00mV 2500.0mV

0.05%+5d

1Gohm

MAX.250VAC or DC
(5kohm PTC)

DCV

5.0000V 50.000V 500.00V 1000.0V

About 10.5 Mohm About 10 Mohm About 10 Mohm About 10 Mohm

MAX.1000VDC or
750VAC
(10Mohm)

  • AC mV & V

Function

Range

Accuracy for each freguency band

50~
500Hz

up to  1kHz

up to 5kHz

up to 10kHz

up to 20kHz

AC mV

500.00mV

0.75%+20d for<400mV
1% for >400mV

1.5% for > 50mV

2% for >50mV

3% for 50mV

AC V

5.0000V
50.000V
500.00V
750.0V

0.75%+20d

1%
for ACV Level
more than 10% of
Full Scale Reading

2%
for ACV
>10% of
Full Scale

Not
Specified

Wave Form

Crest Factor

Additional Error

Square
Sine
Triangle
Others
 

1
1.414
1.73
2
3

0.20%
0%
0.30%
0.50%
1.70%

  • AC + DC mV & V

Function

Range

Accuracy for each freguency band

50~
500Hz

up to  1kHz

up to  5kHz

up to  10kHz

up to  20kHz

AC + DC mV

500.00mV

1.5%+50d for<300mV 2.0%         for >300mV

2.5% for > 50mV

3.5% for > 50mV

4.5% for > 50mV

AC + DC V

5.0000V
50.000V
500.00V
750.0V

2.0% + 50d

2.5%
for AC+DCV Level
more than 10% of
Full Scale Reading

3.5%
for AC+DCV
>10% of
Full Scale

Not
Specified

  • DC Current

Function

Shunt

Range

Accuracy

Protection

DC uA
DC mA
DC A

100 ohm
1 ohm
10 Mohm

5000.0 uA
500.00 mA
10.000 A

0.2% + 5d

1A/600V Fast Fuse&
22V M.O.Varistor

0.5% + 5d

15A/600V Fast Fuse

  • AC Current

Function

Shunt

Range

Accuracy

Protection

AC uA
AC mA
AC A

5000.0 uA
500.00 mA
10.000 A

50Hz to 1kHz

0.75%+20d
for<40000count
1% for>40000count

1A/600V Fast Fuse&
22V M.O.Varistor

15A/600V Fast Fuse

  • AC + DC Current

Function

Range

Frequency

Accuracy

Protection

AC + DC uA
AC + DC mA
AC A

5000.0 uA
500.00 mA
10.000 A

DC to 1kHz

1.5% + 50d
for<30000count
2% for>30000count

1A/600V Fast Fuse&
22V M.O.Varistor

15A/600V Fast Fuse

  • Resistance

Range

Resol.

Accuracy

Open / Short

Protection

50.00 OHM

0.01 ohm

1.0% + 20d

2.5V / 1.4mA

250VAC or DC
(500ohm PTC)

5.0000 K
50.000 K
500.00 K
50000 M

0.1 ohm
1 ohm
10 ohm
100 ohm

0.2% + 5d

2.5V / 1.7mA
2.5V / 240uA
2.5V / 25uA
2.5V / 2.5uA

50.000 M

1k ohm

1.0% + 10d

2.5V / 250nA

  • LV ohm (Low Voltage OHM)

Range

Resol.

Accuracy

Open / Short

Protection

5.000 k
50.00 k
500.0 k
5.000 M

1 ohm
10 ohm
100 ohm
1k ohm

0.2% + 5d  

0.25V / 170uA
0.25V / 24uA
0.25V / 2.5uA
0.25V / 250uA

250VAC or DC
(500 ohm PTC)

50.00 M

10k ohm

1.0% + 10d

0.25V / 25nA

  • nS

Range

Resol.

Accuracy

Open / Short

Protection

500.0 nS

0.1 nS

3.0% + 5d

2.5V / 25nA

250VAC or DC

  • Continuity

Range

Input

Response

Accuracy

Protection

5 kohm
 
50 kohm
 
500 kohm
 
5 Mohm
 

short
open
short
open
short
open
short
open

Buzz On
Buzz Off
Buzz On
Buzz Off
Buzz On
Buzz Off
Buzz On
Buzz Off

< 5 ohm
> 15 ohm
< 50 ohm
> 100 ohm
< 500 ohm
> 1 kohm
< 5 kohm
> 10 kohm

250VAC or DC
(500ohm PTC)

  • Zener Diode

Range

Accuracy

Test Current

Open circuit V

Protection

15.000V

5% + 10d

1mA

15 to 22V

250VAC or DC
(500 ohm PTC)

  • Capacitance

Range

Resol.

Accuracy

Testin Time

Protection

5.000 nF
50.00 nF
500.0 nF
5.000 uF
50.00 uF
500.0 uF
5.000 mF

1 pF
10 pF
100 pF
1 nF
10 nF
100 nF
1 uF

5% + 10d
2% + 10d
 
2% + 5d
 
3% + 10d
 

1.1 Sec

250 VAC or DC
(500 ohm PTC)

Variable
(1.1 ~ 10 Sec.)

  • Frequency

Range

Resol.

Accuracy

Input Impedance

Protection

50.00 Hz
500.0 Hz
50.000 kHz
500.00 kHz
5.0000 kHz

0.01 Hz
0.1 Hz
1 Hz
10 Hz
100 Hz

0.01% + 5d

10Mohm // 1nF

500VAC or DC
(1kV 1nF Capacitor
Parallel with 10M
1/2W Resistor)

  • Puls Width

Range

Resol.

Accuracy

Input Impedance

Protection

200.00 mS

10 uSec.

0.1% + 5d

10Mohm // 1nF

500VAC or DC

  • Duty cycle

Range

Resol.

Accuracy

Pulse width

Protection

100.00%

0.01%

0.1% + 5d

0.1ms ~ 200ms

500VAC or DC

  • Temperature

Range

Resol.

Range

Accuracy

Protection

'C

 
'F
'K

0.1'C

 
0.1'F
0.1'F

-20 ~ 0'C
0 ~ 150'C
150 ~ 1200'C

 ±3'C
 ±2'C 
 ±2%  

Max. 500VAC or
DC for V terminal,
1A 600V Fast Fuse
for mA terminal

'F = 32 +(9/5 x 'C)
'K = 'C + 273.1

  • Peak +/- detection

Function

Accuracy

Remark

DCmA
DCV
DcuA
DCmA
DCA

Less than 10% for peak with
>5ms of duration and with
>10% of full scale of each range

 

  • AD conversion
    50,000 Count Resolution
    0.05% Basic Accuracy and 50ms of Sampling Time
  • Pre-Requisition for Normal Use
    1 Year Calibration Cycle 23+5'C of Ambient Temperature <90%RH of Humidity & Non-condensing
  • Accuracy Expr.
    ±(% of Reading + Least Significant Digits)
  • Bar-graph
    Binary 17 Bit Expression of Mian Digit Decimal
  • Polarity Indication
    - For Negative, None for Positive
  • LCD Update Rate
    0.5 second typ. Exception for the Below
    Functions 1~10 second for Hz,PW,capacitance
  • Memory Locations
    10 Memories for 10 Measured Data
  • Power Source :
    9V Battery , Carbon Zinc or Alkaline Neda 1604,6F22,006P
  • Auto Power Off
    15 min. Typ. After Last Operation of Key or switch
  • Auto Back Light Off
    Automatically Turns Off Typically 30 Sec. after the Back Light Turned on by Pressing the Light Key.
  • Power Consumption & Battery Life
    1) Normal Operating : 7mA typ. which is
         Equivalent to 36 Hours of Continuous
         Use with general Zinc-Carbon Type 9V
         Battery
    2) Auto Power Off : 70uA typ.
         (Equiv. to 5 Month)
    3) Power Switch Off : 7uA Max.
         (Equiv. to 4Year)
  • Safety
    CE Requirement
  • Temp. Coefficient
    <0.1 x Accuracy Spec. / 'C for 0~18'C and
    28~40'C(32~64.4'F and 82.4~104'F)
  • Operating Temp.
    0~40'C(32~104'F)
  • Relative Humidity
    <90%RH at 0~28'C(32~82.4'F),Non Condensing
    <80%RH at 28~40'C(82.4104'F),Non Condensing
  • Storage Temp.
    -20~60'C at ,70%RH & Non Condense
  • Basic Accessary
    Test Leads, User Manual, Gift Box, Holster
    RS232C CD for win95 & win3.1
    RS232C Adaptor Cable
  • Option Accessary
    Temp. Adaptor (608TA)
    Temp. Probe(K)
  • Low Battery Detect
    Battery Indicator on LCD Appearse at Below 6±1V
  • Physical Size (H x W x D)
    Mian Unit(Net) : 203.5 x 94.0 x 43.2mm
    With Holster : 216.5 x 104.5 x 58.0mm
  • Physical Weight
    Main Unit(Net) : 417g(14.7 oz)
    With Holster : 557g(1 lb 3.6oz)

 


  Back to HUNG CHANG Products or
  Back to top of the page

For further information, please
Contact Us

Request a Quote

 


GIGA Electronics Ltd. - Innovative Electronic Solutions

At GIGA Electronics Ltd. our main aim is to offer to our customers the best and most effective solution to meet their
test & measurement needs through exceptional quality and reliability of our products and services.
We take a personal approach to each customer according to his individual needs.